Appendix A
591
Manual Changes
Manual Changes
A. Manual
Changes
The line measurement is performed, and an asterisk appears at the
[ ]Line1 1-2
button and
then a check mark [v] also appears on the
Line1
button.
Step 12.
Press the
Line2
button (6 in Figure A-6).
Step 13.
Connect the Line 2 standard between test ports 1 and 2 and press the
[ ]Line2 1-2
button.
The line measurement is performed, and an asterisk appears at the
[ ]Line2 1-2
button and
then a check mark [v] also appears on the
Line2
button.
NOTE
When the frequency ranges of the line and the match overlap, the data of the line is used in
the overlapped frequency area. When the frequency ranges of multiple lines overlap, the
data of the line measured later is used in the overlapped frequency areas.
NOTE
Isolation calibration measurement is not available.
Step 14.
When all of the necessary calibration measurements are completed, press the
Update Cal
Coef
button (7 in Figure A-6). The calibration coefficient is calculated and the error
correction is turned ON. In the calibration property display (ON/OFF with
[Cal]
-
Property
), the state is indicated by “F” (same as full n-port SOLT calibration).
4. Closing macro
Press the
X
button (8 in Figure A-6) to exit from the macro.
Multiport TRL/LRM calibration operating procedure
1. Selecting Test Port
Select test ports (1 in Figure A-7). As shown in Figure A-7, 3- and
4-port TRL/LRM calibration can be performed with any combination of test ports.
NOTE
For more information on starting a VBA macro, see “Operating Procedure” on page 586.
Figure A-7
Selecting test port
NOTE
The TRL/LRM calibration VBA macro automatically detects the ENA’s active channel,
and calibration is performed for the active channel. The macro does not perform calibration
for the ENA’s inactive channels.
2. Defining Calibration Kit
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...