Appendix D
Softkey Functions
Average Menu
Average Menu
Key Operation
Function
SCPI Command
Displays softkeys for setting averaging options.
Averaging Restart
Resets the counter and restarts from “1”.
:SENS{1-16}:AVER:CLE
Avg Factor
Sets the averaging factor. The averaging factor must be defined as an
integer from 1 to 999.
:SENS{1-16}:AVER:COUN
Averaging
Enables or disables averaging execution.
:SENS{1-16}:AVER
Ave Trigger
Enables or disables averaging trigger.
:TRIG:AVER
Smo Aperture
Sets the smoothing aperture using a percentage against the sweep span
:CALC{1-16}:SMO:APER
Smoothing
Enables or disables the smoothing function. When enabled, the letters
“Smo” will appear in the trace status area.
:CALC{1-16}:SMO:STAT
IF Bandwidth
Sets the IF bandwidth. For more about the effective range of the IF
bandwidth, refer to “Specifications and Reference Data”. If an input value
is out of the range, it will automatically be modified to the closest value in
the effective range. A narrow IF bandwidth slows down the sweep speed,
but improves the S/N ratio.
:SENS{1-16}:BAND
or
:SENS{1-16}:BWID
Return
Returns to the “E5070B/E5071B Menu (Top Menu)” on page 646.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...