Appendix D
663
Softkey Functions
Calibration Menu
D. Sof
tke
y
Fu
nctions
Calibration Menu
Key Operation
Function
SCPI Command
Displays softkeys for setting and executing calibrations.
Correction
Enables or disables error correction.
:SENS{1-16}:CORR:STAT
Calibrate
Displays softkeys for selecting calibration options.
Response (Open)
Displays softkeys for selecting options for response calibration using the OPEN standard.
Select Port
Displays softkeys for selecting a test port.
1
Selects test port 1.
:SENS{1-16}:CORR:COLL:METH:OPEN 1
2
Selects test port 2.
:SENS{1-16}:CORR:COLL:METH:OPEN 2
3
*1
Selects test port 3.
:SENS{1-16}:CORR:COLL:METH:OPEN 3
4
*2
Selects test port 4.
:SENS{1-16}:CORR:COLL:METH:OPEN 4
Cancel
Returns to the softkey display screen one level higher.
Open
Executes a response calibration using the OPEN standard on the
test port selected in the
Select Port
menu. This calibration is
effective for eliminating response tracking errors.
:SENS{1-16}:CORR:COLL:OPEN
Load (Optional)
Executes an isolation calibration using the LOAD standard on the
test port selected in the
Select Port
menu. This calibration is
effective for eliminating directivity errors.
:SENS{1-16}:CORR:COLL:LOAD
Done
Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Cancel
Displays softkeys for canceling a calibration.
OK
Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel
Returns to the softkey display screen one level higher.
Return
Returns to the softkey display screen one level higher.
Response (Short)
Displays softkeys for selecting options for a response calibration using the SHORT standard.
Select Port
Displays softkeys for selecting a test port.
1
Selects test port 1.
:SENS{1-16}:CORR:COLL:METH:SHOR 1
2
Selects test port 2.
:SENS{1-16}:CORR:COLL:METH:SHOR 2
3
Selects test port 3.
:SENS{1-16}:CORR:COLL:METH:SHOR 3
4
Selects test port 4.
:SENS{1-16}:CORR:COLL:METH:SHOR 4
Cancel
Returns to the softkey display screen one level higher.
Short
Executes a SHORT calibration on the test port selected in the
Select
Port
menu. This calibration is effective for eliminating reflection
tracking errors.
:SENS{1-16}:CORR:COLL:SHOR
Load (Optional)
Executes an isolation calibration using the LOAD standard on the
test port selected in the
Select Port
menu. This calibration is
effective for eliminating directivity errors.
:SENS{1-16}:CORR:COLL:LOAD
Done
Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Cancel
Displays softkeys for canceling a calibration.
OK
Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel
Returns to the softkey display screen one level higher.
Return
Returns to the softkey display screen one level higher.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...