Appendix D
725
Softkey Functions
Trigger Menu
D. Sof
tke
y
Fu
nctions
Trigger Menu
Key Operation
Function
SCPI Command
Displays following softkeys for setting the trigger. Once the trigger mode is set, measurements are executed
according to the trigger mode even when the channel is no longer on display due to reduction of the number of
channels to be displayed from the “Display Menu” on page 688.
Hold
Sets the active channel trigger mode to “hold”. A trigger sent from the
trigger source to that channel will not prompt a sweep.
:INIT{1-16}:CONT
Single
Sets the active channel trigger mode to “single”. A trigger sent from the
trigger source to that channel prompts a single sweep and then the sweep
mode changes to “hold.”
:INIT{1-16}
Continuous
Sets the active channel trigger mode to “continuous”. Each trigger sent
from the trigger source to that channel prompts a single sweep.
:INIT{1-16}:CONT
Hold All Channels
Sets all channel trigger modes to “hold”.
None
Continuous Disp
Channels
Sets all displayed channel trigger modes to “continuous”. See “Setting
channel display (layout of channel windows)” on page 62 for details about
displayed channels.
None
Trigger Source
Displays softkeys for selecting the trigger source.
Internal
Selects the “internal” trigger source of the analyzer. The analyzer will
generate a series of trigger signals. A trigger, once generated, is sent to all
channels in order no matter how many channels are displayed.
*1.It depends on the settings of trigger scope.
*1
:TRIG:SOUR INT
External
Selects the “external” trigger source. A signal input to the external trigger
input terminal (BNC(f) connector) on the rear panel will be used as the
trigger source. A trigger, once generated, is sent to all channels in order no
matter how many channels are displayed.
:TRIG:SOUR EXT
Manual
Sets the trigger source to “manual”. Pressing the
Trigger
key in the Trigger
Menu generates a trigger. A trigger, once generated, is sent to all channels
in order no matter how many channels are displayed.
:TRIG:SOUR MAN
Bus
Sets the analyzer trigger source to “bus”. A trigger is generated by sending
a trigger command through the GPIB or a LAN. A trigger, once generated,
is sent to all channels in order no matter how many channels are
displayed.
:TRIG:SOUR BUS
Cancel
Returns to the softkey display screen one level higher.
Trigger Event
Turns ON/OFF the point tirgger function.
:TRIG:POIN
Trigger Scope
Sets the effective range to be triggered.
:TRIG:SCOP
Restart
Aborts a sweep.
:ABOR
Trigger
When the trigger source is set to “manual”, generates a trigger.
:TRIG
Ext Trig Delay
When the low-latency external trigger is set to “ON”, sets the trigger delay
time.
:TRIG:SEQ:EXT:DEL
Low Latency
Enables/disables the low-latency external trigger.
:TRIG:SEQ:EXT:LLAT
Return
Returns to the “E5070B/E5071B Menu (Top Menu)” on page 646.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...