24
Chapter 2
Installation
Starting the E5071C
This operation turns ON the power, and the E5071C starts the self-test.
Step 4.
Confirm that the self-test indicates normal operation.
Normal operation is confirmed by the self-test if no error message appears.
Turning the Power OFF
Step 1.
Use either of the following methods to turn the power OFF.
•
Press the standby switch (
) in the lower-left part of the front panel (now in the
pressed down (
) position) to put it in the popped up (
) position.
•
Click Turn off Computer from Windows Start menu
These operations will start the E5071C shutdown process (required software and hardware
processes for turning the power off), and the power will turn OFF after a few seconds.
NOTE
Under normal circumstances, always press the standby switch (
), or click Turn off
Computer from Windows Start menu, to execute the E5071C shutdown process.
Never cut
off the power supply directly by disconnecting the power cable plug from the rear
panel of the unit.
If the power supply is cut off directly by disconnecting the power cable plug from the
instrument or the AC outlet, the shutdown process is not carried out and there is a risk of
damage to the E5071C’s software or hardware.
Figure 2-7
Line switch (Always ON) and power cable receptacle
e5071cqe1003
EXT TRIG
IN
EXT TRIG
OUT
GPIB
115 V
350 VA Max
LINE
-
230 V
50/60 Hz
REF IN
REF OUT
AUX IN 2
AUX IN 1
WARNING: No operator serviceable
parts inside. Refer servicing to qualified
personnel.
4
LR95111C
N10149
ICES/NMB-001
ISM 1-A
24 BIT I/O
USB
Serial Label
Model# Label
REF OVEN
(10 MHz)
1 V, 10 V
10 MHz
PRINTER
(PARALLEL)
VIDEO
RESERVED
RESERVED
ETHERNET
2
1
4
3
I/O PANEL
Windows Label
Switch must remain
ON while operating.
NOTE
FUSE 0.5 A
PORT 1
PORT 2
PORT 3
PORT 4
BIAS INPUT 35 VDC Max
Power Cable Receptacle
Line Switch: Always ON ( | )
Summary of Contents for E5071C ENA Series
Page 6: ......
Page 14: ...8 Contents ...
Page 36: ...30 Chapter2 Installation Calibration of the Touch Screen ...