Chapter 3
Troubleshooting
Troubleshooting during Startup
Troubleshooting during Startup
When you encounter the problem during start up, see Table 3-1. System recovery saves
most of problems.
Table 3-1
Troubleshooting during startup
Symptom
Solution
Turning on (|) the standby switch does not
start up the system.
Confirm that the power cable is properly
plugged in.
Confirm that the line switch on the rear
panel is turned on
The system starts up, but it automatically
shuts down immediately.
Execute the system recovery.
The system starts up, but it enters the
service mode (The instrument status bar
in the lower right part of the screen
displays SVC in red).
The measurement screen appears after
startup, but the power-on test is failed,
with Error Message 241 appearing against
a red background in the instrument
message/warning area in the lower left
part of the screen.