Appendix B
359
System Recovery
Calibration of the Touch Screen
B.
Sy
stem
Recovery
Calibration of the Touch Screen
After E5071C measurement screen appears, you have to calibrate the touch screen. Follow
the procedure described below to calibrate the touch screen.
Step 1.
Press
.
Step 2.
Press
Service Menu
.
Step 3.
Press
Test Menu
.
Step 4.
Press
Adjust Touch Screen
.
The touch screen calibration screen (Figure B-7) appears.
Figure B-7
Touch Panel Calibration Screen
e5071cse1078
Step 5.
Touch the x mark on the upper left with your finger. The mark x appears also on the upper
right, lower left, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the
touch screen calibration.
NOTE
With no operation on the touch screen calibration screen for a preset time, it automatically
closes and the previous measurement screen reappears.
Summary of Contents for E5071C
Page 14: ...14 Contents ...
Page 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Page 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Page 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Page 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Page 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Page 364: ...364 AppendixC Firmware Update Firmware Update ...
Page 365: ...365 D Power Requirement D Power Requirement ...