382
Appendix E
Messages
User characterization not found in module
This error occurs also when a port not existing on this model is specified in a command
specifying a port number as an index. Such commands are
CALC:FSIM:SEND:DEEM:PORTn:xxxx, CALC:FSIM:SEND:PMC:PORTn:xxxx,
CALC:FSIM:SEND:ZCON:PORTn:Z0:R, and SENS:CORR:EXT:PORTn:TIME; they
include PORTn as a part.
34
User characterization not found in module
This error occurs when the selected user profile is not detected in the ECal memory, while
reading it from the ECal module, written by the user definition ECal.
V
30
Valid Ecal module not found
This error occurs when the number of ports of the ECal module connected is less than the
necessary number of ports. This error occurs, for example, when a 4-port Cal executing
command, SENS:CORR:COLL:ECAL:SOLT4, is executed with a 2-port ECal module
connected. This error is not generated by front key operations.
71
Valid power meter not found
Valid power meter is not connected. (refer to “Power Calibration” as described in E5071C
Online Help)
This error occurs when the E5071C and the power meter is not connected with USB/GPIB
interface, the GPIB addresses do not match between the power meter and the E5071C’s
power meter, or the power meter is turned off, during the measurement for power
calibration data or for that in scalar-mixer calibration.
It also occurs when inoperable power meter is connected.
Ensure the connection of the power meter, GPIB addresses and power supply.
72
Valid power sensor not found
No valid sensor is connected to the power meter.
This error occurs when no power sensor is connected to the specified channel of the power
meter, during the measurement for power calibration data or for that in scalar-mixer
calibration.
Ensure the connection of the power sensor.
75
Valid signal generator not found
No valid external signal source is connected.
This error occurs when the specified type of external signal source is not connected to the
specified GPIB address, though the control of the external signal source is enabled
(SENS:OFFS:LOC:STAT ON and SENS:OFFS:LOC:CONT ON).
It also occurs when USB/GPIB interface is set but not connected.
Ensure the external signal source and the connection of USB/GPIB interface.
Summary of Contents for E5071C
Page 14: ...14 Contents ...
Page 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Page 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Page 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Page 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Page 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Page 364: ...364 AppendixC Firmware Update Firmware Update ...
Page 365: ...365 D Power Requirement D Power Requirement ...