272
Chapter 6
Replacement Procedure
Removable Hard Disk Drive Replacement
Restore Backup Files onto the New Hard Disk Drive
The following procedure shows how to write the system correction data stored in the
backup memory to the newly mounted hard disk.
Step 1.
Turn on the E5071C.
Step 1.
Press
.
Step 2.
Click
Load Project...
in the softkey menu.
Step 3.
Select
RestoreSysCorFile.vba
from the
D:\Agilent\Service
folder and then press the
Open
button.
Step 1.
Press
.
Step 2.
The RestoreSysCorFile dialog box appears.
Step 3.
Click
OK
.
NOTE
Once the program has terminated, restart the E5071C to enable the system correction data.
Calibration of the Touch Screen
When you have replaced the HDD on a E5071C equipped, you have to calibrate the touch
screen. Follow the procedure described below to calibrate the touch screen.
Step 1.
Press
.
Step 2.
Press
Service Menu
.
Step 3.
Press
Test Menu
.
Step 4.
Press
Adjust Touch Screen
.
The touch screen calibration screen (Figure 6-3) appears.
Summary of Contents for E5071C
Page 14: ...14 Contents ...
Page 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Page 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Page 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Page 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Page 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Page 364: ...364 AppendixC Firmware Update Firmware Update ...
Page 365: ...365 D Power Requirement D Power Requirement ...