Chapter 2
75
Performance Test
E5071C Performance Test Record
2. Pe
rfor
ma
nce T
est
CW Frequency: 8.5 GHz (relative to 0 dBm reference)
Option 280, 285, 480, 485
Power level
[dBm]
Test limit [dB]
Test result [dB]
Measurement
uncertainty [dB]
-20
±
0.75
±
0.24
-17.5
±
0.75
±
0.24
-15
±
0.75
±
0.24
-12.5
±
0.75
±
0.24
-10
±
0.75
±
0.24
-7.5
±
0.75
±
0.24
-5
±
0.75
±
0.24
-2.5
±
0.75
±
0.24
2.5
±
0.75
±
0.23
5
±
0.75
±
0.23
7
±
0.75
±
0.23
Trace Noise Test
Option 230
Direction: S11 (Magnitude)
Frequency
Range[Hz]
IF bandwidth
[Hz]
Test limit
[dB rms]
Test result
[dB rms]
Measurement
uncertainty [dB]
9 k - 30 k
3 k
< 0.004
0.00009
30 k - 10 M
3 k
< 0.003
0.00009
10 M - 3 G
70 k
< 0.004
0.00033
Direction: S11 (Phase)
8
±
0.75
±
0.23
Power level
[dBm]
Test limit [dB]
Test result [dB]
Measurement
uncertainty [dB]
Frequency
Range[Hz]
IF bandwidth
[Hz]
Test limit
[deg rms]
Test result
[deg rms]
Measurement
Uncertaintiy [deg]
9 k - 30 k
3 k
< 0.035
0.00077
30 k - 10 M
3 k
< 0.02
0.0006
Summary of Contents for E5071C
Page 14: ...14 Contents ...
Page 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Page 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Page 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Page 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Page 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Page 364: ...364 AppendixC Firmware Update Firmware Update ...
Page 365: ...365 D Power Requirement D Power Requirement ...