26
Chapter 2
Performance Tests
7. LOAD MATCH, INSERTION LOSS, ISOLATION for SWITCH 5 (SPDT, PORT
5, 5A and 5B)
7. LOAD MATCH, INSERTION LOSS, ISOLATION for
SWITCH 5 (SPDT, PORT 5, 5A and 5B)
Description
This test checks the Load Match, Insertion Loss and Isolation for ports 5COM, 5A and 5B
of the E5092A. These parameters checked at 50 MHz through 20 GHz with the N5230A
Network Analyzer.
Specification
@ 23
°
C
±
5
°
C (Warming up over 90 minutes)
Load Match (Selected port)
Load Match (Unselected port)
Load Match (Common port)
SPDT Switch
Specification
50 MHz to 2 GHz
17 dB
2 GHz to 4 GHz
11 dB
4 GHz to 8 GHz
8 dB
8 GHz to 10 GHz
7 dB
10 GHz to 18 GHz
4 dB
18 GHz to 20 GHz
4 dB
SPDT Switch
Specification
50 MHz to 3 GHz
17 dB
3 GHz to 10 GHz
11 dB
10 GHz to 16 GHz
8 dB
16 GHz to 18 GHz
6 dB
18 GHz to 20 GHz
4 dB
SPDT Switch
Specification
50 MHz to 2 GHz
16 dB
2 GHz to 4 GHz
11 dB
4 GHz to 8 GHz
8 dB
8 GHz to 10 GHz
7 dB