Chapter 2
133
Performance Tests
E5092A Performance Test Record
2. Pe
rfor
ma
nce T
e
st
s
From Selected Port To Unselected Port
Control Line Maximum Current Test
From Port
- To Port
(Selected
Port)
Frequency
Range [Hz]
Test Limit
[dB]
Test Result [dB]
Measurement
Uncertainty
[dB]
10A - 10B
(10A)
50M to 500M
< -65.0
±
0.35
500M to 1G
< -80.0
±
0.49
1G to 2G
< -85.0
±
0.62
2G to 6G
< -90.0
±
0.71
6G to 10G
< -85.0
±
0.71
10G to 18G
< -75.0
±
0.44
18G to 20G
< -65.0
±
0.36
10B - 10A
(10B)
50M to 500M
< -65.0
±
0.35
500M to 1G
< -80.0
±
0.49
1G to 2G
< -85.0
±
0.62
2G to 6G
< -90.0
±
0.71
6G to 10G
< -85.0
±
0.71
10G to 18G
< -75.0
±
0.44
18G to 20G
< -65.0
±
0.36
Control Line
Group
Lower Test
Limit [mA]
Test Result [mA]
Measurement
Uncertainty [mA]
A
≥
50
±
0.008856
B
≥
50
±
0.008856
C
≥
0.5
±
0.000046
D
≥
0.5
±
0.000046
Summary of Contents for E5092A
Page 10: ...4 ...
Page 140: ...134 Chapter2 Performance Tests E5092A Performance Test Record ...
Page 152: ...146 Chapter3 Troubleshooting Performance Test Failure Troubleshooting ...
Page 180: ...174 Chapter4 Replaceable Parts Replaceable Parts List for E5092A 020 ...
Page 205: ...199 B Power Requirement A Power Requirement ...