Chapter 2
45
Performance Tests
E5092A Performance Test Record
2. Pe
rfor
ma
nce T
e
st
s
E5092A Performance Test Record
Agilent Technologies E5092A Configurable Multiport Test Set
DC Source Output Voltage Accuracy Test
DC Source Maximum Current Test
Serial Number:
Option:
Temperature:
°
C
Test Date:
Humidity:
%RH
Tested by:
Control
Line
Group
DC
Source
Voltage
[V]
Test
Limit [V]
Test Result [V]
Measurement
Uncertainty
[V]
TAR
A
1.000
±
0.030
±
0.000015
2058
A
5.000
±
0.150
±
0.000054
2803
B
1.000
±
0.030
±
0.000015
2058
B
5.000
±
0.150
±
0.000054
2803
C
1.000
±
0.030
±
0.000015
2058
C
5.000
±
0.150
±
0.000054
2803
D
1.000
±
0.030
±
0.000015
2058
D
5.000
±
0.150
±
0.000054
2803
Control Line
Group
Lower Test Limit
[mA]
Test Limit[mA]
Measurement
Uncertainty [mA]
A
≥
150
±
0.0582
B
≥
150
±
0.0582
C
≥
150
±
0.0582
D
≥
150
±
0.0582