Agilent E5270 TIS User’s Guide, Edition 1
2-53
TIS Function Reference
set_pbias
Figure 2-5
Staircase Sweep with Pulsed Bias Measurements
To start staircase sweep with pulsed bias measurement, execute the set_iv function,
set_pbias function (this function), and then sweep_iv function. The sweep source
ignores the hold time and delay time defined by the set_iv function, and changes the
sweep output value in the timing set by this function. See Figure 2-5.
NOTE
Functions that disable the set_pbias settings
If the set_iv function is executed after this function, the set_pbias settings will be
cleared, and the set_iv settings will be effective.
If program contains this function and the set_piv function, the set_pbias settings will
not be effective, and the set_piv settings will be effective.
If the disable_port function is executed for the same
port
set to this function, the
set_pbias settings will be cleared.
Also the set_pbias settings will be cleared by the init_system function that initializes
the Agilent E5270.