2-12
Agilent E5270 TIS User’s Guide, Edition 1
TIS Function Reference
Arguments
Table 2-6
Current Source Setup Parameters
range
value
(Output Range)
Output Value
a
a. Setting resolution is determined by actual output range used to force the output current.
Setting
Resolution
Maximum Compliance
MPSMU
HPSMU
0 < |
range
|
£
1 nA
(1 nA)
0
£
|I|
£
1 nA
50 fA
±100 V
±200 V
1 nA <
|
range
|
£
10 nA
(10 nA)
0
£
|I|
£
10 nA
500 fA
±100 V
±200 V
10 nA <
|
range
|
£
100 nA
(100 nA)
0
£
|I|
£
100 nA
5 pA
±100 V
±200 V
100 nA < |
range
|
£
1
m
A
(1
m
A)
0
£
|I|
£
1
m
A
50 pA
±100 V
±200 V
1
m
A < |
range
|
£
10
m
A
(10
m
A)
0
£
|I|
£
10
m
A
500 pA
±100 V
±200 V
10
m
A <
|
range
|
£
100
m
A
(100
m
A)
0
£
|I|
£
100
m
A
5 nA
±100 V
±200 V
100
m
A < |
range
|
£
1 mA
(1 mA)
0
£
|I|
£
1 mA
50 nA
±100 V
±200 V
1 mA < |
range
|
£
10 mA
(10 mA)
0
£
|I|
£
10 mA
500 nA
±100 V
±200 V
10 mA < |
range
|
£
100 mA
(100 mA)
0
£
|I|
£
20 mA
5
m
A
±100 V
±200 V
20 mA < |I|
£
50 mA
±40 V
50 mA < |I|
£
100 mA
±20 V
±100 V
100 mA <
|
range
|
£
200 mA
(200 mA for MPSMU)
0
£
|I|
£
20 mA
10
m
A
±100 V
-
20 mA < |I|
£
50 mA
±40 V
-
50 mA < |I|
£
200 mA
±20 V
-
100 mA <
|
range
|
£
1 A
(1 A for HPSMU)
0
£
|I|
£
50 mA
50
m
A
-
±200 V
50 mA < |I|
£
125 mA
-
±100 V
125 mA < |I|
£
500 mA
-
±40 V
500 mA <
|I|
£
1 A
-
±20 V