Agilent Technologies E5500 Phase Noise Measurement System
8-19
Residual Measurement Fundamentals
The Calibration Options
Synthesized Residual
Measurement using
Beatnote Cal
This calibration option requires two synthesizers of which one must be
tunable such that a beatnote can be acquired. For the system to calibrate, the
beatnote frequency must be within the following ranges.
Procedure
1. Connect circuit as per
, and tighten all
connections.
Figure 8-12
Synthesized Residual Measurement using Beatnote Cal
2. Offset the carrier frequency of one synthesizer to produce a beatnote for
cal.
Table 8-5
Beatnote Frequency Ranges
Carrier Frequency
Beatnote Frequency
Range
<500 kHz
10 Hz to 10 kHz
<5 MHz
10 Hz to 100 kHz
<50 MHz
10 Hz to 1 MHz
<250 MHz
10 Hz to 10 MHz
>250 MHz
10 Hz to 50 MHz
or 1/2 the frequency range of the configured analyzer, or
whichever is lower.