Agilent Technologies E5500 Phase Noise Measurement System
11-17
FM Discriminator Measurement Examples
FM Discriminator Measurement using Double-Sided Spur Calibration
5
Test Set Tab
• The test set parameters do not apply to this measurement
example.
6
Dowconverter Tab
• The downconverter parameters do not apply to this
measurement example.
7
Graph Tab
• Title
• Graph Type
• X Scale Minimum
• X Scale Maximum
• Y Scale Minimum
• Y Scale Maximum
• Normalize trace data to a:
• Scale trace data to a new
carrier frequency of:
• Shift trace data DOWN by:
• Trace Smoothing Amount
• Power present at input of DUT
• FM Discrim - 50 ns dly - 1.027GHz, +19 dBm out, VCO,DSS
• Single-sideband Noise (dBc/Hz)
• 10 Hz
• 100 E + 6 Hz
• 10 dBc/Hz
• - 190 dBc/Hz
• 1 Hz bandwidth
• 1 times the current carrier frequency
• 0 dB
• 0
• 0 dB
Step
Parameters
Data