13-8
Agilent Technologies E5500 Phase Noise Measurement System
AM Noise Measurement Examples
AM Noise using an Agilent/HP 70420A Option 001
Figure 13-1
Connect Diagram for the AM Noise Measurement
4. Refer to the following system connect diagram example for more
information about system interconnections:
• Internal AM Detector
0 to +20 dBm
• Downconverters:
Agilent/HP 70422A
+5 to +15 dBm
Agilent/HP 70427A
0 to +30 dBm
CAUTION:
To prevent damage to the Agilent/HP 70420A test set’s hardware components,
the input signal must not be applied to the test set’s signal input connector until
the input attenuator (Option 001) has been correctly set by the phase noise
software, which will occur at the connection diagram.
Characteristics:
Input Impedance
50 ohm Nominal
AM Noise
dc coupled to 50 ohm load
Table 13-2
Agilent/HP 70420A Test Set Signal Input Limits and Characteristics