Agilent Technologies E5500 Phase Noise Measurement System
A-9
Connector Care and Preventive Maintenance
Removing and Reinstalling Instruments
•
Turn the silver nut clockwise by hand until it is snug, then tighten with
an 8 inch-lb torque wrench (part number 8720-1765). This wrench may
be ordered from Agilent Technologies.
Bent Semirigid Cables
Semirigid cables are not intended to be bent outside of the factory. An
accidental bend that is slight or gradual may be straightened carefully by
hand. Semirigid cables that are crimped will affect system performance, and
must be replaced. Do not attempt to straighten a crimped semirigid cable, its
performance will not be restored.
Other Multipin
Connectors
There are other multipin connectors in the system (Agilent/HP MSIB, for
example). These are sometimes held in place by a pair of screws.