Agilent Technologies E5500 Phase Noise Measurement System
Expanding Your Measurement Experience
Testing the Agilent/HP 8644B Internal/External 10 MHz
Selecting a Reference
Source
1. From the Define menu, choose Measurement; then choose the Block
Diagram tab from the Define Measurement window.
2. From the Reference Source pull-down list, select HP-8644.
3. When you have completed these operations, click the Close button
.
Agilent/HP 8644B
FM Deviation
0
10
600
Compute
Other Signal
Generator
DCFM Calibrated for
±1V
FM Deviation
0
10
R
in
Compute
Other User VCO
Source
Estimated within a
factor of 2
–10 to
+10
1 E + 6
Measure
VCO Source
Carrier
Freq.
Tuning Constant
(Hz/V)
Center
Voltage
(V)
Voltage Tuning
Range (
±
V)
Input
Resistance
(
Ω)
Tuning
Calibration
Method