5-40
Agilent Technologies E5500 Phase Noise Measurement System
Expanding Your Measurement Experience
Testing the Agilent/HP 8644B Internal/External 10 MHz
CAUTION
The Agilent/HP 70420A test set’s signal input is subject to the following
limits and characteristics:
Table 5-7
Agilent/HP 70420A Test Set Signal Input Limits and Characteristics
Limits
Frequency
50 kHz to 26.5 GHz
Maximum Signal Input Power
+30 dBm
At Attenuator Output, Operating Level
Range:
• RF Phase Detectors
0 to +23 dBm
• Microwave Phase Detectors
0 to +5 dBm
• Internal AM Detector
0 to +20 dBm
• Downconverters:
Agilent/HP 70422A
0 to +30 dBm
Agilent/HP 70427A
+5 to +15 dBm
CAUTION:
To prevent damage to the Agilent/HP 70420A Test Set’s hardware components,
the input signal must not be applied to the test set’s signal input connector until
the input attenuator (Option 001) has been set by the phase noise software,
which will occur at the connection diagram.
Characteristics:
Input Impedance
50 ohm Nominal
AM Noise
dc coupled to 50 ohm load