40
E5505A Installation Guide
4
Measurement Software
Table 11
Parameter Data for the Agilent N5500A Confidence Test Example
Step
Parameters
Data
1
Type and Range Tab
•
Measurement Type
•
Start Frequency
•
Stop Frequency
•
Minimum Number of Averages
FFT Quality
Swept Quality
•
Baseband noise (using a test set)
•
10 Hz
•
100E + 6 Hz (determined by analyzer
used)
•
4
•
Fast
•
Fast
2
Cal Tab
Gain preceding noise input
•
0 dB
3
Block Diagram Tab
Noise Source
•
Test set Noise input
4
Test Set Tab
Input Attenuation
LNA Low Pass Filter
•
LNA Gain
•
DC Block
•
PLL Integrator Attenuation
•
Ignore out-of-lock conditions
•
Pulsed Carrier
•
0 dB
•
20 MHz (auto checked)
•
Auto Gain (minimum auto gain –14 dB)
•
Not checked
•
0 dBm
•
Not checked
•
Not checked
5
Graph Tab
•
Title
•
Graph Type
•
X Scale Minimum
•
X Scale Maximum
•
Y Scale Minimum
•
Y Scale Maximum
•
Normalize trace data to a:
•
Scale trace data to a new carrier
frequency of
•
Shift trace data by
•
Trace Smoothing Amount
•
Power present at input of DUT
•
E5500 Confidence Test
•
Base band noise (dBV/Hz)
•
10 Hz
•
100 E + 6 Hz
•
0 dBc/Hz
•
–200 dBc/Hz
•
1 Hz bandwidth
•
1 times the current carrier frequency
•
0 dB
•
0
•
0 dBm
Summary of Contents for E5505A
Page 18: ...18 E5505A Installation Guide 1 General Information ...
Page 22: ...22 E5505A Installation Guide 2 Flat Panel Display ...
Page 30: ...30 E5505A Installation Guide 3 System Interconnections ...
Page 84: ...84 E5505A Installation Guide 6 Preventive Maintenance ...
Page 96: ...96 E5505A Installation Guide A Service Support and Safety Information ...