78
E5505A Installation Guide
6
Preventive Maintenance
When reconnecting this type of cable:
•
Carefully insert the male connector center pin into the female
connector. (Make sure the cable is aligned with the instrument
connector properly before joining them.)
•
Turn the silver nut clockwise by hand until it is snug, then tighten with
an 8 inch- lb torque wrench (part number 8720- 1765).
Bent Semirigid Cables
Semirigid cables are not intended to be bent outside of the factory. An
accidental bend that is slight or gradual may be straightened carefully by
hand. Semirigid cables that are crimped will affect system performance
and must be replaced. Do not attempt to straighten a crimped semirigid
cable.