Troubleshooting the E5810B4
E5810B User’s Guide
99
viopen fails or I/O operation fails: Out of
resources
The E5810B was not able to open another session or perform the
operation because it is out of resources. You may need to have a client
terminate its connection or you may need to reboot the E5810B to close
all client connections to reclaim resources. Rebooting will disconnect all
other users accessing the E5810B.
I/O operation fails: No connection
The connection to the E5810B failed because:
•
The network connection is down or the E5810B is down and network
timeout occurred.
•
The E5810B was powered-off and then powered-on, or the E5810B
reboot was executed, either of which terminates any open client
connections.
I/O operation fails: Timeout
There are several possible causes:
•
The client may be attempting a transaction to a non-existent or
powered-off device. Check the address used, as well as the status of
the device.
•
The network connection or the E5810B has gone down since the
previous I/O operation.
•
The I/O timeout specified is not long enough for the transaction to
complete. You may need to verify that the E5810B I/O timeout
configuration value is long enough if the client specified a timeout
value of infinity.
•
The E5810B is busy performing an I/O operation for another client or
another client has the device/interface locked. Refer to
“E5810B
appears to be hung”
for more information.
Table 4-6
Common runtime messages (continued)
Message
Description
Summary of Contents for E5810B
Page 1: ...Agilent Technologies Agilent E5810B LAN GPIB USB Gateway User s Guide ...
Page 8: ...VIII E5810B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...XIV E5810B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...XVIII E5810B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 156: ...Glossary V 138 E5810B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...