E5810B Description
1
E5810B Software/Firmware Descriptions
E5810B User’s Guide
11
The E5810B and the connected instruments can be placed
anywhere on the network (rather than with a particular
controller or a server computer system). This includes
networks which span different geographic locations, such as
when networked computer systems are located at several
different sites.
GPIB, USB, and RS- 232 instruments can be located where
they are most convenient. You can also have more than one
E5810B on a network, providing clusters of GPIB, USB, and
RS- 232 instruments at different locations.
Software/firmware architecture
Client system architecture
Figure 1-6
Client system architecture
Client System (PC)
Application
Agilent VISA
SICL
LAN client
TCP
IP
LAN interface
Server (E5810B)
LAN Server
TCP
IP
LAN
interface
IO
driver
Instrument
Instrument firmware
GPIB/USB/RS- 232 bus