E5810B Web Access3
E5810B User’s Guide
63
Checking for connected instruments
If you do not see one or more of your connected
instruments in the list, try these steps:
1
Check the cables.
Make sure that all the instruments are connected via
GPIB cable, USB cable, or RS- 232 cable to the E5810B
GPIB port, USB port, or RS- 232 port respectively.
2
Turn on the instrument and select
Find
at the bottom of
the Instruments Connected column.
3
Check Hardware.
A broken instrument may short out the cable and cause
the instrument to show a failure. For GPIB instruments, if
you are experiencing problems, try connecting only one
GPIB instrument to the E5810B at a time. Possible causes
of the problem may be broken cable, loose connectors, two
or more instruments at the same GPIB address, an
instrument may be broken, or the E5810B may be broken.
4
Reboot the E5810B.
After checking the hardware connections, reboot the
E5810B to do a self- test for the E5810B hardware.