Chapter 5
207
Language Reference
CALCulate:EMI:SLISt Subsection
CURRent - current signal.
TOEnd - from the current signal to the end of the list.
DUPLicates - all duplicate signals.
LOWer- all lower amplitude duplicates.
COMPlement - complement signals.
Factory Preset
and *RST:
Not affected by preset
Front Panel
Access:
MEASURE, More, Signal List, Signal Marking, Mark Signal
Position Cursor to Signal List
:CALCulate:EMI:SLISt:SELect FIRSt|LAST|NEXT|PREVious|
<integer>
Position cursor to signal in list.
Factory Preset
and *RST:
Not affected by preset
Remarks:
Once this function is defined (
Signal List On
), the cursor is
positioned and the selected state is persistent. Persistent means
that it retains the setting previously selected, even through a
power cycle.
This command takes a signal-index value which corresponds to
the position of a signal in the signal list. Valid range is 1 to 2000.
Front Panel
Access:
MEASURE, More, Signal List, Signal List
Specify Sort Key
:CALCulate:EMI:SLISt:SORT
FREQuency|PEAK|QPEak|AVERage|LLINE1|LLINE2
ASCending|DESending
:CALCulate:EMI:SLISt:SORT FREQ, ASC?
Specifies the sort key and order for signal list sorting. Both the sort and order are
required. The format is comma separated.
Factory Preset
and *RST:
Not affected by preset
Front Panel
Access:
MEASURE, More, Signal List, Sort Signals, By Freq
MEASURE, More, Signal List, Sort Signals, By Pk Ampl
MEASURE, More, Signal List, Sort Signals, By QP Ampl
MEASURE, More, Signal List, Sort Signals, By
∆
LL1
MEASURE, More, Signal List, Sort Signals, By
∆
LL2