228
Chapter 5
Language Reference
CALCulate:NTData Subsection
CALCulate:NTData Subsection
Normalize the Trace Data
:CALCulate:NTData[:STATe] OFF|ON|0|1
:CALCulate:NTData[:STATe]?
One sweep of trace data is copied to trace 3 (firmware version greater then
A.03.03,
NRML
in firmware version less than or equal to A.03.03), which is used as
the reference trace. Then for all subsequent trace sweeps, display trace 1
=
data
collected into trace 1 – data in trace 3 (firmware version greater than A.03.03,
NRML in firmware version less than or equal to A.03.03).
Front Panel
Access:
View/Trace, Normalize, Normalize On Off