4-48
Service Guide E8364-90001
Troubleshooting
PNA Series Microwave Network Analyzers
Measurement System Troubleshooting
E8362A, E8363A, E8364A
For Port 2 measurements, set the network analyzer for an S
22
measurement with a CW
frequency of 1 GHz.
Perform the following checks in the order presented.
Checking the A22 Switch Splitter and A25 and A26 Test Port Couplers
1. Locate the following semirigid cable at the A22 switch splitter:
Standard . . . . . . . . . . . . . . . . . . . . . . . . W5 for Port 1 or W6 for Port 2
Option UNL . . . . . . . . . . . . . . . . . . . . . . W51 for Port 1 or W52 for Port 2
Option 014 . . . . . . . . . . . . . . . . . . . . . . . W61 for Port 1 or W62 for Port 2
Option UNL/014 . . . . . . . . . . . . . . . . . . W51 for Port 1 or W52 for Port 2
2. Using a 5/16-inch torque wrench, disconnect the semirigid cable at the A22 switch
splitter.
3. Connect the spectrum analyzer to the connector. Set the spectrum analyzer to measure
a signal at 1 GHz.
4. If the 1 GHz signal is not present, replace the A22 switch splitter. Refer to
and Replacing the A22 Switch Splitter” on page 7-40
5. If the 1 GHz signal is present:
Standard . . . . . . . . . replace the appropriate test port coupler. Refer to
Replacing the A25 and A26 Test Port Couplers” on page 7-46
Option 014 . . . . . . . . replace the appropriate test port coupler. Refer to
Replacing the A25 and A26 Test Port Couplers” on page 7-46
Option UNL . . . . . . . reconnect the semirigid cable and go to
A37 60-dB Step Attenuators (Option UNL)” on page 4-48
Option UNL/014 . . . reconnect the semirigid cable and go to
A37 60-dB Step Attenuators (Option UNL)” on page 4-48
Checking the A36 and A37 60-dB Step Attenuators (Option UNL)
1. Locate the following semirigid cable at the appropriate step attenuator:
Option UNL . . . . . . . . . . . . . . . . . . . . . . W53 for Port 1 or W54 for Port 2
Option UNL/014 . . . . . . . . . . . . . . . . . . W81 for Port 1 or W82 for Port 2
2. Using a 5/16-inch torque wrench, disconnect the semirigid cable at the step attenuator.
3. Connect the spectrum analyzer to the step attenuator connector. Set the spectrum
analyzer to measure a signal at 1 GHz.
4. If the 1 GHz signal is not present, replace the appropriate 60-dB step attenuator. Refer
“Removing and Replacing the A36 and A37 60-dB Step Attenuators (Option UNL)” on
5. If the 1 GHz signal is present, reconnect the semirigid cable and go to
Summary of Contents for E8362A
Page 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Page 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Page 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Page 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Page 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Page 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Page 287: ...Service Guide E8364 90001 A 1 A Error Terms ...
Page 303: ...Service Guide E8364 90001 B 1 B Option Enable Utility ...
Page 309: ...Service Guide E8364 90001 C 1 C Firmware Upgrades ...
Page 313: ...Service Guide E8364 90001 D 1 D Operating System Recovery ...