2-4
Service Guide E8364-90001
General Product Information
PNA Series Microwave Network Analyzers
Analyzer Options, Upgrades, and Accessories
E8362A, E8363A, E8364A
(
+
30 dBm) can be applied to the device under test (DUT). Additionally, there is an external
reference input that allows the external amplifier’s frequency response and drift to be
ratioed out. There are also internal step attenuators between the coupler and the receivers
to prevent receiver overload.
For high dynamic range measurements, front panel jumpers are moved to reverse the
signal path through one of the couplers, allowing for a 15 dB improvement in transmitted
signal sensitivity in one direction only. Jumpers are installed on both ports allowing the
user to choose a measurement in either the forward or reverse direction.
Option 099, Firmware Upgrade
This upgrade provides the latest revision of firmware for the PNA series network
analyzers on CD-ROM. This firmware is user installable. The installation requires a USB
CD-ROM drive (see
) or an external computer connected to a
Local Area Network (LAN).
This firmware is also available on the Internet at the Agilent website:
http://www.agilent.com/find/pna
. (Select your analyzer’s model number in this website to
view available upgrades.)
Option UNL, Source Attenuators and Bias Tees
This option adds a 60-dB step attenuator and a bias tee between the switch splitter and
each of the ports. The step attenuators are used to adjust the power level to the device
under test (DUT) without changing the power in the reference path. The bias tees provide
a means of biasing active devices under test. DC bias for the bias tees is provided through
two rear-panel BNC connectors. These inputs are fused for protection.
CD-RW Drive
This accessory is an external read/write CD drive with a USB cable.
This accessory can be ordered as model number N4688A.
USB Hub
This accessory is a 4-port USB hub for connecting additional USB peripheral devices.
This accessory can be ordered as model number N4689A.
Option AVK, Printed User’s and Programmer’s Guide
This option adds a printed copy of the user’s and programmer’s documentation that is
embedded in the analyzer.
This option is not available as an upgrade. It can be ordered as part number:
English . . . . . . . . . . . . . . . . . . . . . . .E8356-90028
Summary of Contents for E8362A
Page 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Page 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Page 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Page 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Page 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Page 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Page 287: ...Service Guide E8364 90001 A 1 A Error Terms ...
Page 303: ...Service Guide E8364 90001 B 1 B Option Enable Utility ...
Page 309: ...Service Guide E8364 90001 C 1 C Firmware Upgrades ...
Page 313: ...Service Guide E8364 90001 D 1 D Operating System Recovery ...