PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362A, E8363A, E8364A
Adjustments
Procedure
1. Connect the equipment as shown in
. Connect a GPIB cable between the
network analyzer and the power meter.
Figure 3-19 Setup 1 for the Receiver Calibration Adjustment
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
Receiver Calibration
.
3. Ensure the GPIB settings are correct.
4. Click
Calibrate
, and then follow the instructions as they are displayed.
Figure 3-20 Setup 2 for the Receiver Calibration Adjustment