Service Guide E8364-90001
A- 11
PNA Series Microwave Network Analyzers
Error Terms
E8362A, E8363A, E8364A
Error Term Data
Source Match (E
SF
and E
SR
)
E
SF
and E
SR
are the forward and reverse uncorrected source match terms of the driven
port. They are obtained by measuring the reflection (S
11
, S
22
) of an open, and a short that
are connected directly to the ports. Source match is a measure of the match of the coupler,
as well as the match between all components from the source to the output port.
The measurements most affected by source match errors are reflection and transmission
measurements of highly reflective DUTs.
Typical Cause of Failure
The
calibration kit open or shor
t is the most common cause of source match specification
failure.
If the open or short performance has been independently verified, then suspect the
analyzer
switch splitter, step attenuator, or coupler
.
To troubleshoot, refer to
“Checking the Signal Separation Group” on page 4-46
.
Figure A-4 Typical E
SF
/E
SR
Without Cables
Summary of Contents for E8362A
Page 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Page 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Page 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Page 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Page 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Page 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Page 287: ...Service Guide E8364 90001 A 1 A Error Terms ...
Page 303: ...Service Guide E8364 90001 B 1 B Option Enable Utility ...
Page 309: ...Service Guide E8364 90001 C 1 C Firmware Upgrades ...
Page 313: ...Service Guide E8364 90001 D 1 D Operating System Recovery ...