Service Guide E8364-90001
3- 25
PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362A, E8363A, E8364A
System Verification
System Verification Procedure
1. Connect a cable to Port 1 on the analyzer as shown in
. If you desire printed
test outputs, connect a printer to the analyzer. For the printer, ensure that the correct
driver is loaded and the printer is defined as the default printer. Refer to the embedded
help in the analyzer for printer setup. Let the analyzer warm up for at least 90 minutes.
Figure 3-11 System Verification Test Setup
2. Insert the verification kit disk into the analyzer disk drive.
3. On the
System
menu, point to
Service
, and then click
System Verification
. The
System
Verification
dialog box is displayed; refer to
Figure 3-12 System Verification Dialog Box
Summary of Contents for E8362A