7-22
Service Guide E8364-90026
Repair and Replacement Procedures
PNA Series Microwave Network Analyzers
Removing and Replacing the A12 Source 20
E8362B, E8363B, E8364B
Removing and Replacing the A12 Source 20
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16-inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
for this procedure.
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to
“Removing the Covers” on page 7-6
.
3. Remove the A13 (Option 080) frequency offset receiver assembly, if present. Refer to
“Removing and Replacing the A13 Frequency Offset Receiver Assembly” on page 7-24
4. With a 5/16-inch torque wrench, disconnect the semirigid cable (item
①
) from the
bottom of the A12 source 20.
CAUTION
To avoid damage to the semirigid cable (item
①
), pull down on the cable from
the bottom of the analyzer to move it away while sliding out the
A12 source 20. The cable has some spring to it and the RF pin could be
damaged by the A12 source 20.
5. Disengage the A12 source 20 from the A11 connector and slide out of the analyzer.
Observe the
CAUTION
above.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
Summary of Contents for E8362B
Page 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Page 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Page 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Page 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Page 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Page 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...