7-50
Service Guide E8364-90026
Repair and Replacement Procedures
PNA Series Microwave Network Analyzers
Removing and Replacing the A27, A28, A29, and A30 First Converters (Mixers) and Mixer Bias Board E8362B, E8363B, E8364B
Removing and Replacing the A27, A28, A29, and A30
First Converters (Mixers) and Mixer Bias Board
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16-inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
for this procedure.
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
“Removing the Covers” on page 7-6
. Position the
analyzer bottom side up and raise the receiver deck as shown. Refer to
Removing the Receiver Deck” on page 7-8
if necessary.
3. With a 5/16-inch torque wrench, disconnect the eight semirigid cables (item
①
) and the
four flexible cables (item
②
).
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing
of the cables is necessary to remove the assembly. Do not over-bend them.
4. With a T-10 TORX driver, remove the four mounting screws (item
③
).
5. Remove the mixer bracket and disconnect the ribbon cable (item
④
). Observe the
CAUTION
above.
6.
If replacing a first converter (mixer):
a. Remove the defective mixer from the bracket by removing the attachment screws.
b. Install the new mixer onto the bracket using the attachment screws from the mixer
that was removed. Be careful to properly align the DC feed-through pins.
If replacing the mixer bias board:
a. Remove all four mixers from the bracket by removing the attachment screws.
b. Remove the mixer bias board by removing the attachment screws.
c. Install the new mixer bias board onto the bracket by
loosely
installing the
attachment screws from old the mixer bias board. Do not tighten them at this time.
d. Reinstall the four mixers using their original attachment screws. Be careful to
properly align the DC feed-through pins.
e. Using a T-10 TORX driver, tighten the mixer bias board attachment screws.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
Summary of Contents for E8362B
Page 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Page 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Page 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Page 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Page 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Page 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...