Service Guide E8364-90026
8-17
PNA Series Microwave Network Analyzers
General Purpose Maintenance Procedures
E8362B, E8363B, E8364B
Option Enable Utility
Repairing and Recovering Option Data
License, option, and model number data are stored in an EEPROM on the frequency
reference board and written into the
gen.lic
file on the hard disk drive. If the data stored
in either of these items is lost, it can be recovered from the other item.
NOTE
If the data on both the hard disk drive and the frequency reference board is
lost, it can not be recovered. Contact Agilent for assistance. Refer to
“Contacting Agilent” on page 2-13
.
Recovery of Data After Repair
• If the frequency reference board is replaced, use the
Repair
selection to recover data that
has been lost as a result of the repair:
1. Select
Repair
from the
Option Enable
menu bar (see
).
2. Click the
Freq Ref board has been replaced
check box.
3. Click
Begin Repair
. The data is written from the
gen.lic
file into the EEPROM.
• If the hard disk drive is replaced, a new
gen.lic
file is automatically created when the
Network Analyzer
application starts, by retrieving the data from the EEPROM.
Recovery of Data if Option or Model Numbers are Incorrect
If the analyzer option or model numbers are not listed correctly on the analyzer display, in
either the Option Enable Dialog Box or the About Network Analyzer display, you should
regenerate the options license file,
gen.lic
:
1. Exit the
Network Analyzer
application.
2. Remove (or rename) the existing
gen.lic
file:
a. Open
Windows Explorer
and navigate to
C:\Program Files\Agilent\Network
Analyzer
.
b. Delete (or rename)
gen.lic
.
3. Exit the
Windows Explorer
application.
4. Open the
Network Analyzer
application. The application will generate a new
gen.lic
file when it starts.
5. Check the option listing:
a. On the
System
menu, point to
Service
, and then click
Option Enable
, or
b. On the
Help
menu, click
About Network Analyzer
.
NOTE
If the options are still not listed correctly, contact Agilent for assistance. Refer
to
“Contacting Agilent” on page 2-13
.
Summary of Contents for E8362B
Page 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Page 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Page 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Page 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Page 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Page 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...