3-44
Service Guide E8364-90026
Tests and Adjustments
PNA Series Microwave Network Analyzers
Adjustments
E8362B, E8363B, E8364B
8. Reconnect the power and USB cables. Turn on the analyzer.
9. On the
System
menu, point to
Service
,
Adjustments
, and then click
LO Power Adjust
.
10. Ensure the GPIB settings are correct.
11. Click
Begin Adj
,
and then follow the instructions as they are displayed.
12. Turn off the analyzer, disconnect the power cable, reconnect the semi-rigid cables, and
replace the covers.