Service Guide E8364-90026
4-31
PNA Series Microwave Network Analyzers
Troubleshooting
E8362B, E8363B, E8364B
Measurement System Troubleshooting
Verifying the A, B, R1, and R2 Traces (Standard S-Parameter Mode)
NOTE
There is no way to view the frequency offset receiver response (Option 080).
However, some standard S-parameter receiver trace information is helpful in
troubleshooting the frequency offset section of the PNA. It is therefore
recommended that you run this test even if you
only
suspect the frequency
offset section of malfunctioning.
The first step is to verify that the A, B, R1, and R2 traces are present and that they are
approximately level:
• On the
System
menu, point to
Service
,
Utilities
, and then click
Receiver Display
.
• Traces A, B, R1, and R2 are displayed in four separate data windows as shown in
. Identifying discrepancies of the traces in these windows can help you to isolate the
faulty assembly.
Figure 4-9 Typical Four Channel Display