4-58
Service Guide E8364-90026
Troubleshooting
PNA Series Microwave Network Analyzers
Measurement System Troubleshooting
E8362B, E8363B, E8364B
• The spectrum analyzer should measure a signal at 1 GHz.
5. If the measured signal is not present, replace A46 switch. Refer to
Replacing the A45 and A46 Reference Channel Transfer Switches (Option 081)” on
page 7-70
6. If the measured signal is present, reconnect W60.
7. From the
Channel
menu, click
Test Set
. In the
Test Set
dialog box, click
Internal: bypass R1
loop
. Click
OK
.
8. Remove semirigid cable W96 from A45 switch input.
9. Connect the spectrum analyzer to the W96 cable.
10.The spectrum analyzer should measure a signal at 1 GHz.
11. If the measured signal is not present replace, A46 switch. Refer to
Replacing the A45 and A46 Reference Channel Transfer Switches (Option 081)” on
page 7-70
12. If the measured signal is present then reconnect W96.
13. Remove W99 from the A45 switch output.
14. Connect the spectrum analyzer to the output of the A45 switch.
15. The spectrum analyzer should measure a signal at 1 GHz.
16. If the measured signal is not present replace A45 switch. Refer to
Replacing the A45 and A46 Reference Channel Transfer Switches (Option 081)” on
page 7-70
17. If the instrument passes all the tests to this point but you still have a problem with
your PNA, then refer to
“Contacting Agilent” on page 2-13
.
Summary of Contents for E8362B
Page 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Page 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Page 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Page 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Page 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Page 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...