5-18
Service Guide E8364-90026
Theory of Operation
PNA Series Microwave Network Analyzers
Signal Separation Group Operation
E8362B, E8363B, E8364B
A36 and A37 60-dB Source Step Attenuators (Option UNL)
The 60-dB step attenuators provide coarse power control for the test port signals. They are
electro-mechanical step attenuators that provide 0 to 60 dB of attenuation in 10-dB steps.
These attenuators are controlled by the A15 CPU board. They adjust the power level to the
DUT without changing the level of the incident power in the reference path.
A38 and A39 Bias Tees (Option UNL)
The A38 and A39 bias tees provide a means of biasing active devices with an external dc
voltage connected to the rear-panel DC BIAS CONNECT ports. The dc voltage is applied
directly to the center conductor of the test port connectors. A blocking capacitor ensures
that the dc bias current goes only to the DUT, and not back into the source. Similarly, an
inductor in the bias path prevents RF from being imposed on the external dc supply.
A43 and A44 35-dB Receiver Step Attenuators (Option 016)
These 35-dB step attenuators provide coarse power control for amplified input. They are
electro-mechanical step attenuators that provide 0 to 35 dB of attenuation in 5-dB steps.
These attenuators are controlled by the A15 CPU board.
Option 014 Configurable Test Set
The Option 014 analyzer allows you to measure devices with higher power and higher
dynamic range limits than the standard analyzer. The theory of operation is the same as
for the standard analyzer except that there are six front panel SMA jumpers.
As shown in
, these jumpers allow access to the test port and
reference signals between:
• the A22 switch splitter and the A25 test port 1 coupler. (For Option UNL, this is
between the A36 step attenuator and the A38 bias tee.)
• the A22 switch splitter and the A26 test port 2 coupler, (For Option UNL, this is
between the A37 step attenuator and the A39 bias tee.)
• the A25 test port 1 coupler and the A27 receiver A first converter
• the A26 test port 2 coupler and the A30 receiver B first converter
• the A23 detector (for E8363B and E8364B) or fixed attenuator (for E8362B) and the
A28 receiver R1 first converter
• the A24 detector (for E8363B and E8364B) or fixed attenuator (for E8362B) and the
A29 receiver R2 first converter
Summary of Contents for E8362B
Page 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Page 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Page 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Page 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Page 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Page 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...