Service Guide E8364-90026
2-3
PNA Series Microwave Network Analyzers
General Product Information
E8362B, E8363B, E8364B
Maintenance
Maintenance
WARNING
To prevent electrical shock, disconnect the analyzer from the mains
source before cleaning. Use a dry cloth or one slightly dampened
with water to clean the external case parts. Do not attempt to clean
internally.
Physical Maintenance
Clean the cabinet, including the front panel, using a dry or slightly damp cloth only.
Electrical Maintenance
“Contacting Agilent” on page 2-13
Chapter 3, “Tests and Adjustments.”
Analyzer Options Available
Options as Upgrades
The options described in this section can be ordered as upgrades. Refer to
Upgrades Available” on page 2-7
.
Individual upgrade kits are also available and can be ordered by part number. Refer to
“E8362B Upgrade Kits” on page 6-81
“E8363B, E8364B Upgrade Kits” on page 6-81
for a complete list of available kits.
Option 010, Time Domain
An Option 010 analyzer can display the time domain response of a network or test device
by calculating the inverse Fourier transform of the frequency domain response. This
calculation allows the Option 010 analyzer to show the response of a test device as a
function of time or distance. In time domain, the magnitude and location of a discontinuity
and the characteristics of individual transmission paths of a network or test device can be
determined. Time domain operation retains all the accuracy of active error correction.
Option 014, Configurable Test Set
An Option 014 analyzer can be configured to measure high-power devices and devices for
high dynamic range.
For a high-power measurement, external amplifiers and high power attenuators or
isolators can be added to complete the test setup. Test port output power up to 1 Watt
(
+
30 dBm) can be applied to the device under test (DUT). Additionally, there is an external
reference input that allows the external amplifier’s frequency response and drift to be
ratioed out.
Summary of Contents for E8362B
Page 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Page 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Page 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Page 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Page 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Page 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...