Service Guide E8364-90026
8-21
PNA Series Microwave Network Analyzers
General Purpose Maintenance Procedures
E8362B, E8363B, E8364B
Correction Constants
Correction Constants
The analyzer stores many correction constants in non-volatile EEPROM memory. These
constants enable the analyzer to produce accurate, leveled source signals and receive clean
test signals.
Storing Correction Constants
After performing any adjustment listed on
in this manual, store the correction
constants to a backup file on the analyzer hard disk drive by performing these steps:
• Navigate to the EEPROM Backup Utility, located at:
C:\Program Files\Agilent\Network Analyzer\Service\eebackup.exe
•
Run
the program.
• Click
Backup EEPROM
.
• Click
Exit
when the program has finished.