Service Guide E8364-90026
3-13
PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362B, E8363B, E8364B
Preliminary Checks
The Test Port Cable Checks
A faulty test port cable can cause a failure in the verification test. The following checks are
not required, but are recommended to verify the performance of the test port cable.
•
“Cable Return Loss Check” on page 3-14
•
“Cable Insertion Loss Check” on page 3-15
•
“Cable Magnitude and Phase Stability Check” on page 3-16
•
“Cable Connector Repeatability Check” on page 3-18
Accessories Used in the Test Port Cable Checks
Equipment Type
Models
Used With
Model or
Part Number
Alternate Model
or Part Number
Calibration kit, 3.5 mm
E8362B
85052B
85052D
RF cable, 3.5 mm (f) to 3.5 mm (f)
E8362B
85131C
85131E
Calibration kit, 2.4 mm
E8363B, E8364B
85056A
85056D
RF cable, 2.4 mm (f) to 2.4 mm (f)
E8363B, E8364B
85133C
85133E