PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362B, E8363B, E8364B
Preliminary Checks
12. For the Port 2 Check, connect the cable (with the load attached) to Port 2 and repeat
steps 2 through 11.
Figure 3-7
Typical Cable Connector Repeatability Response
If the Cable Connector Repeatability Check Fails
1. Clean the cable and devices, and torque connections to specifications (refer to
the Principles of Connector Care” on page 3-5
). Repeat the check.
2. If the check still fails, the cable should be repaired or replaced.