Service Guide E8364-90026
3-25
PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362B, E8363B, E8364B
System Verification
System Verification Procedure
1. Connect a cable to the analyzer as shown in
. If you desire printed test
outputs, connect a printer to the analyzer. For the printer, ensure that the correct driver
is loaded and the printer is defined as the default printer. Refer to the embedded help in
the analyzer for printer setup. Let the analyzer warm up for at least 90 minutes.
Figure 3-11 System Verification Test Setup
2. Insert the verification kit disk into the analyzer disk drive.
3. On the
System
menu, point to
Service
, and then click
System Verification
. The
System
Verification
dialog box is displayed; refer to
Figure 3-12 System Verification Dialog Box
Summary of Contents for E8362B
Page 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Page 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Page 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Page 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Page 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Page 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...