4-38
Service Guide E8364-90038
Troubleshooting
PNA Series Microwave Network Analyzers
Measurement System Troubleshooting
E8362C, E8363C, E8364C
Checking the Signal Separation Group
Before checking the signal separation group, perform the following procedures:
•
“Getting Ready to Test” on page 4-31
Checking the Output Power of the A and B Signals
Using a power meter, you can measure the outputs of the A and B signals from the front panel. The
measurement results will help you isolate a faulty assembly. The outputs of the R1 and R2 channels cannot
be measured because it would necessitate breaking the phase lock loop, causing all of the signals to be lost.
Equipment Used for These Tests
Equipment Setup
1. Before starting these checks, zero and calibrate the power meter. (See the power meter user’s guide for
instructions on setting the calibration factor.)
2. If the
Receiver Display
(
Figure 4-6
) is not on the analyzer screen, perform the following: On the
System
menu, point to
Service, Utilities,
and then click
Receiver Display.
3. Set the sweep speed for a 10 second sweep: On the
Sweep
menu, click
Sweep Time
and set the time to
10.000 seconds in the
Sweep
Time
box.
Checking Port 1 Power (A Signal)
The object of this check is to verify the power of the A signal across the entire frequency range. Perform this
test if there is an observed problem only with the channel A trace. The ten second sweep is slow enough to
allow you to observe the output power on the power meter as the sweep occurs.
1. Connect the power sensor to Port 1.
2. Observe the power reading on the power meter as the sweep occurs on the analyzer.
3. The measured output power on the power meter should be within -12 dBm ±4 dB for E836xB with no
options or -17 dBm ±4 dB for E836xB with options, over the entire frequency range.
•
If the measured power is correct, go to
“Checking the Receiver Group” on page 4-42
.
•
If the measured power is not correct, go to
“Checking the Signal through the Signal Separation Path”
on page 4-39
.
Equipment Type
Models
Used With
Model or
Part Number
Alternate Model or
Part Number
Power meter
E4418B/E441B
E4418A/E4419A
Power sensor
E8362C
E8485A
None
Adapter, 3.5 mm (f) to 3.5 mm (f)
E8362C
83059B
85052-60012
Power sensor, 2.4 mm
E8363C, E8364C
8487A
None
Adapter, 2.4 mm (f) to 2.4 mm (f)
E8363C, E8364C
11900B
85056-60007
Summary of Contents for E8362C
Page 11: ...Service Guide E8364 90038 1 1 1 Safety and Regulatory Information ...
Page 21: ...Service Guide E8364 90038 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90038 3 1 3 Tests and Adjustments ...
Page 81: ...Service Guide E8364 90038 4 1 4 Troubleshooting ...
Page 137: ...Service Guide E8364 90038 5 1 5 Theory of Operation ...
Page 171: ...Service Guide E8364 90038 6 1 6 Replaceable Parts ...
Page 257: ...Service Guide E8364 90038 7 1 7 Repair and Replacement Procedures ...
Page 341: ...Service Guide E8364 90038 8 1 8 General Purpose Maintenance Procedures ...