8-18
Service Guide E8364-90038
General Purpose Maintenance Procedures
PNA Series Microwave Network Analyzers
Option Enable Utility
E8362C, E8363C, E8364C
Installing or Changing a Serial Number
It is necessary to reinstall the instrument serial number if the test set motherboard is replaced.
IMPORTANT
Use extreme care when entering the serial number, as only one attempt is allowed.
1. To change a serial number, select
Change
Serial
from the
Option
Enable
menu bar (see
Figure 8-5 on
page 8-15
). The current serial number is displayed. If no serial number has previously been entered, the
word “NONE” will be displayed.
2. VERY CAREFULLY, type the new serial number into the space provided and then click
Change
.
3. If an error is made in entering the serial number, obtain a clear code from Agilent, enter the clear code in
the space provided, and then click
Clear
. The correct serial number can then be entered.
NOTE
To change an incorrect serial number, a clear-code password is required. Contact Agilent to
obtain the clear-code. Refer to
“Contacting Agilent” on page 2-12
.
Summary of Contents for E8362C
Page 11: ...Service Guide E8364 90038 1 1 1 Safety and Regulatory Information ...
Page 21: ...Service Guide E8364 90038 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90038 3 1 3 Tests and Adjustments ...
Page 81: ...Service Guide E8364 90038 4 1 4 Troubleshooting ...
Page 137: ...Service Guide E8364 90038 5 1 5 Theory of Operation ...
Page 171: ...Service Guide E8364 90038 6 1 6 Replaceable Parts ...
Page 257: ...Service Guide E8364 90038 7 1 7 Repair and Replacement Procedures ...
Page 341: ...Service Guide E8364 90038 8 1 8 General Purpose Maintenance Procedures ...