Service Guide E8364-90038
3- 13
PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362C, E8363C, E836C
Preliminary Checks
Cable Return Loss Check
1. Press
Preset
.
2. Perform a one-port calibration on Port 1,
1-Port Reflection
. Refer to the embedded help in the analyzer if
necessary.
3. Connect the test port cable to Port 1. Connect a broadband load to the other end of the cable. Tighten to
the specified torque for the connector type.
The analyzer now displays the return loss of the cable.
4. From the
Marker
menu, click
Marker Search
. In the
Marker Search
dialog box, in the
Search Type
box,
make sure
Maximum
is selected. Click
Execute
, and then click
OK
.
5. The marker annotation on the screen indicates the worst case return loss. Refer to the cable manual to
see if it meets the return loss specification. For an example of a typical return loss measurement,
see
Figure 3-4
.
Figure 3-4
Typical Cable Return Loss Response
If the Cable Return Loss Check Fails
1. Clean the cable and devices and torque to specification. Repeat the check.
2. If the check still fails, the cable should be repaired or replaced.
Summary of Contents for E8362C
Page 11: ...Service Guide E8364 90038 1 1 1 Safety and Regulatory Information ...
Page 21: ...Service Guide E8364 90038 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90038 3 1 3 Tests and Adjustments ...
Page 81: ...Service Guide E8364 90038 4 1 4 Troubleshooting ...
Page 137: ...Service Guide E8364 90038 5 1 5 Theory of Operation ...
Page 171: ...Service Guide E8364 90038 6 1 6 Replaceable Parts ...
Page 257: ...Service Guide E8364 90038 7 1 7 Repair and Replacement Procedures ...
Page 341: ...Service Guide E8364 90038 8 1 8 General Purpose Maintenance Procedures ...