3-40
Service Guide E8364-90038
Tests and Adjustments
PNA Series Microwave Network Analyzers
Adjustments
E8362C, E8363C, E836C
3.8 GHz PMYO Frequency Adjustment
The 3.8 GHz adjustment is used to adjust the frequency accuracy of the network analyzer’s 3.8 GHz
permanent-magnet YIG oscillator (PMYO) on the A12 source 20.
Equipment used for the Frequency Adjustment at 3.8 GHz
Procedure
1. Connect the equipment as shown in
Figure 3-18
. Connect a GPIB cable between the network analyzer
and the spectrum analyzer.
Figure 3-18
Setup for Adjustment of the 3.8 GHz PMYO
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
3.8 GHz Freq Adjust
.
3. Ensure the GPIB settings are correct.
4. Click
Begin Adj
, and then follow the instructions as they are displayed.
Equipment Type
Models
Used With
Model or
Part Number
Alternate Model
or Part Number
Spectrum analyzer
All
8565E
856xE
a
a. Must be capable of measuring a signal at 3.8 GHz.
RF Cable, 3.5 mm (f) to 3.5 mm (f)
E8362C
85131C
85131E
Adapter, 3. 5 mm (m) to type-N (m)
E8362C
1250-1743
None
RF Cable, 2.4 mm (f) to 2.4 mm (f)
E8363C, E8364C
85133C
85133E
Adapter, 2.4 mm (m) to type-N (m)
E8363C, E8364C
11903A
None
Summary of Contents for E8362C
Page 11: ...Service Guide E8364 90038 1 1 1 Safety and Regulatory Information ...
Page 21: ...Service Guide E8364 90038 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90038 3 1 3 Tests and Adjustments ...
Page 81: ...Service Guide E8364 90038 4 1 4 Troubleshooting ...
Page 137: ...Service Guide E8364 90038 5 1 5 Theory of Operation ...
Page 171: ...Service Guide E8364 90038 6 1 6 Replaceable Parts ...
Page 257: ...Service Guide E8364 90038 7 1 7 Repair and Replacement Procedures ...
Page 341: ...Service Guide E8364 90038 8 1 8 General Purpose Maintenance Procedures ...