Service Guide E8364-90038
3- 45
PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362C, E8363C, E836C
Adjustments
Source Calibration Adjustment
The source calibration is used to adjust your network analyzer for a flat source power across its full
frequency range. The differences between Port 1 and Port 2 are negligible; therefore, a Port 2 adjustment is
not required.
Equipment Used for the Source Calibration Adjustment
Procedure
1. Connect the equipment as shown in
Figure 3-21
. Connect a GPIB cable between the network analyzer
and the power meter.
Figure 3-21
Setup for the Source Calibration Adjustment
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
Source Calibration
.
3. Ensure the GPIB settings are correct.
4. Click
Calibrate
,
and then follow the instructions as they are displayed.
Equipment Type
Models
Used With
Model or
Part Number
Alternate Model or
Part Number
Power meter
All
E4418B/E4419B
E4418A/E4419A
Power sensor, 3.5 mm
E8362C
E4413A
8485A
Adapter, 3.5 mm (f) to 3.5 mm (f)
E8362C
83059B
85052-60012
Power sensor, 2.4 mm
E8363C, E8364C
8487A
None
Adapter, 2.4 mm (f) to 2.4 mm (f)
E8363C, E8364C
11900B
85056-60007
Summary of Contents for E8362C
Page 11: ...Service Guide E8364 90038 1 1 1 Safety and Regulatory Information ...
Page 21: ...Service Guide E8364 90038 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90038 3 1 3 Tests and Adjustments ...
Page 81: ...Service Guide E8364 90038 4 1 4 Troubleshooting ...
Page 137: ...Service Guide E8364 90038 5 1 5 Theory of Operation ...
Page 171: ...Service Guide E8364 90038 6 1 6 Replaceable Parts ...
Page 257: ...Service Guide E8364 90038 7 1 7 Repair and Replacement Procedures ...
Page 341: ...Service Guide E8364 90038 8 1 8 General Purpose Maintenance Procedures ...